Model M3000

M3000 Specs
The M3000 is a compact, bench-top system having a larger
chamber to accommodate multiple probes, data loggers and
small chart recorders. The control probes are calibrated and
adjusted by an NVLAP Accredited laboratory using NIST
traceable methodology. Accuracy of the system is ±1.0%RH
and ±0.2ºC temperature. Chamber stability is ±0.2%RH and
and ±0.1ºC. System operating range is 10-92%RH @ 23ºC
and an overall range of 10º-50ºC. Response time for 50%RH
step typically 3-5 minutes after temperature equilibrium. An
RS-232 port is provided with software/cable for automatic
ramp/soak cycles for linearization or broad range validations.
|