ALow Cost, Fast Response RH Generator/Calibrator with Digital Control
The Model 2000SP is a low cost, high accuracy alternative to the 2-pressure or chilled-mirror humidity calibrations. The Model 2000SP is a state-of-the-art calibration system using digital control for fast response in order to maintain chamber stability. Chamber RH and temperature uniformity is maintained by mixing the internal chamber environment at a high rate of air flow.
The response time for chamber RH equilibrium is typically within 3-5 minutes, for a 50%RH step change, after reaching temperature equilibrium. The RH/temperature range of operation meets most requirements for full range calibration, 5°-60°C and 3-97%RH
Control probes of the Model 2000SP are calibrated by an NVLAP Accredited laboratory using methodology traceable to NIST. Provision is made to use a chilled-mirror instrument (optional) as the reference to monitor chamber conditions and thereby generate the RH indicated value.
The Model 2000SP can satisfy your calibration needs without a high cost
System Features and Advantages
Digital control for better stability of RH and temperature
Chamber stabilization in less than 7 minutes over 50%RH step @ 23°C
Sample ports enables use of the of RH and temperature chilled-mirror as a reference
Recommended calibration intervalfor chamber control probe 12 months
Lightweight and transportable for in-plant or laboratory calibration of probes
Chamber doors designed to accept probes from all manufacturers